Design of integrating sensors based on non-invasive blood sugar measuring

WANG Di-ya,CHEN Zhen-cheng,ZHU Jian-ming,JIN Xing-liang
DOI: https://doi.org/10.3969/j.issn.1000-9787.2009.12.024
2009-01-01
Abstract:Based on the method of conservation of energy metabolism(COEM),an integrator of sensors is designed to measure the blood sugar level non-invasively,which size is 60 mm×30 mm×85.5 mm.Including the base,the blood flow measuring region of the finger surface,the metabolism rate measuring region of the part of finger,the oxygen saturation infrared measuring region of the part of finger,the link region of circuit and the signal in-out region.The integrator is integrated by the temperature sensors,the humidity sensors,the radiation sensors,the dual emitter photodetectors(660 nm/905 nm)at some implementary conditions.Blood sugar level is achieved non-invasively by the integrator.The results of 20 preliminary clinical experimentations suggest that the correlative coefficient is up to 0.807 between the integrator and automatic biochemical analyzer(AUTOLAB18),which shows strong correlation and good coherence.The average relative error is 7.0 %,and the boundary of average relative different is less than 0.5 through the statistical error analysis,which shows the measuremental results is effective and accurate.