Analytical Method of Total Mercury in Contaminated Soils Using Atomic Fluorescence Spectrophotometer

XUE Ruimin,WANG Ronghui,ZHU Xuemei,ZHANG Zengqiang,LIU Feng,HE Jie,YU Hongjin
DOI: https://doi.org/10.3969/j.issn.1004-1389.2012.06.038
2012-01-01
Abstract:The preparation treatment and the measurement method were developed to determine the total mercury in contaminated soils.The standard soil(GBW07405-GSS5) was used,and different parameters were tried and selected including the quantity of the soil,digestion solution volume,digestion time,and analytical instruments.The selected preparation parameters were as follows: 8 mL of 1∶1 aqua regia solution,0.5 g soil samples,18 h digestion in room temperature,and then digestion 60 min in the boiled water bath.The atomic spectrophotometer(AFS-3000) was used to determine the mercury in the digestion solutions.The detection limit of total mercury was 0.005 μg/L,relative standard deviation was 2.7%,the recovery ranged from 94.4 % to 107%.The total mercury of Hg-contaminated soils from different contaminated sites were analyzed using this measurement method,the relative standard deviation ranged from 0.24% to 0.98%,and the statistical results showed that there was no significance among triplicate samples,indicating the method could provide reliable and accurate data.