Electricity Chromatographic Performance Characterization of Tentacle-Type Polymeric Open-Tubular Column

ZHANG Xiao-Wei,YANG Juan,LIN Xu-Cong,XIE Zeng-Hong
DOI: https://doi.org/10.3969/j.issn.1004-8138.2011.06.012
2011-01-01
Abstract:The tentacle-type polymeric open-tubular column that was modified by branched polyethyleneimine(PEI) was prepared by covalently bonded method,and the electrochromatography performance of which was detailedly investigated.The electroosmotic flow(EOF) strength of open-tubular column decreased along with increase of ACN and buffer salt concentration in mobile phase.When pH value of mobile phase was less than 7.5,the open-tubular column got the reversed EOF,and the days,daytime and weeks RSDs for EOF were in the range of 0.8%—2.6%,and between the columns and columns RSD was 1.6%(n=4),while batch and batch RSD was 4.8%(n=3),that showed the good repeatability.The favorable separation of proteins was successfully obtained without obvious peak tailing.
What problem does this paper attempt to address?