Effects of Miniaturization of Cardiopulmonary Bypass Circuits on Blood Transfusion Requirement in Neonatal and Infant Open Heart Surgery

Zhong Hui,He Chunlan,Wang Xiuhua,Chen Zhanggen,Jia Bing
DOI: https://doi.org/10.3969/j.issn.1674-1242.2010.04.007
2010-01-01
Progress in Biomedical Engineering
Abstract:Objectives We conducted a retrospective analysis of the effect of stepwise miniaturization of the cardiopulmonary bypass(CPB) circuit on homologous blood requirement during neonatal and infant open heart surgery.Methods All patients(n=4822) weighing less than 10Kg undergoing open heart surgery between 2000 and 2010 were divided into three groups: group 1(n=1265),Dedico 901/902 /Medtronic oxygenator+Stockert II/Jostra HL20 heart-lung machine;group 2(n=2534),Rx-05 oxygenator+Stockert II/Jostra HL20 heart-lung machine;group 3(n=1023),Rx-05 oxygenator+ Jostra HL30 heart-lung machine.Amount of priming volume,red blood cell use and hematocrit were compared.Results Priming volume in group 3 and 2 was significantly less compared with group 1(group 1,718±15ml;group 2,458±15ml;group 3,334±8ml,P<0.001).The hematocrit values during the perioperative did not differ among the groups;however there was a significant difference in the amount of RBC used in the priming(group 1,313±5 ml;group 2,205±5ml;group 3,167±8ml,P<0.001),as well as during CPB(group 1,117±8ml;group 2,88±7ml;group 3,60±8ml,P<0.001). Conculsion By downsizing the oxygenator and reducing the circuit tubing length,we succeed to reduce the priming volume and subsequently the total use of blood and it is a proper step towards transfusion-free cardiac surgery.