Non-Contact Measurement of Heartbeat Signal Via Doppler Radar Using Adaptive Filtering

ZHANG Hua,LU Guo-hua,JING Xi-jing,WANG Hua,WU Hao,JIAO Teng,ZHANG Yang,YU Xiao,ZHANG Xiao,ZHANG Peng-fei,XUE Hui-jun,WANG Jian-qi
DOI: https://doi.org/10.3969/j.issn.1003-8868.2012.01.004
2012-01-01
Abstract:Objective To separate the heartbeat signal from the mixsignal of the respiration and heartbeat,which was measured via contact-free biological radar.Methods A heartbeat signal separate method was proposed by using recursive-least square(RLS) algorithm based adaptive noise canceler.Results RLS algorithm based adaptive noise canceler could isolate the heartbeat signal,and there was a strong correlation between the heart rate and heart rate obtained from the ECG signal(γ2 = 0.95,P 0.000 1).Conclusion The heartbeat signal separation method present possesses good practicality,which is expected to achieve the real-time separation of respiration and heartbeat signal.