Potentiometric Sensor Based on Molecularly Imprinted Polymer for Determination of Melamine in Milk

Rongning Liang,Ruiming Zhang,Wei Qin
DOI: https://doi.org/10.1016/j.snb.2009.05.024
2009-01-01
Abstract:A polymeric membrane ion-selective electrode for determination of melamine is described in this paper. It is based on a molecularly imprinted polymer (MIP) for selective recognition, which can be synthesized by using melamine as a template molecule, methacrylic acid as a functional monomer and ethylene glycol dimethacrylate as a cross-linking agent. The membrane electrode shows near-Nernstian response (54 mV/decade) to the protonated melamine over the concentration range of 5.0 × 10−6 to 1.0 × 10−2 mol L−1. The electrode exhibits a short response time of ∼16 s and can be stable for more than 2 months. Combined with flow analysis system, the potentiometric sensor has been successfully applied to the determination of melamine in milk samples. Interference from high concentrations of ions co-existing in milk samples such as K+ and Na+ can be effectively eliminated by on-line introduction of anion- and cation-exchanger tandem columns placed upstream, while melamine existing as neutral molecules in milk of pH 6.7 can flow through the ion-exchanger columns and be measured downstream by the proposed electrode in an acetate buffer solution of pH 3.7.
What problem does this paper attempt to address?