Quantification of Urinary Β_2 Microglobulin with Piezoelectric Quarts Crystal Biosensor

LUO Yang,CHEN Ming,ZHANG Bo,ZHAO Meng,HUANG Jun-fu,ZHANG Xue,FU Wei-ling
DOI: https://doi.org/10.3321/j.issn:1005-4529.2008.03.008
2008-01-01
Abstract:OBJECTIVE To discuss and establish a highly effective method to quantify β2 microglobulin in the urine by immunosensor microarray. METHODS Common protein A (SPA) was used to immobilize the antibody on the micro-array. Then detect the concentration of antigen conjugated on the quarts crystal microbalance (QCM) to compare the difference of the sensitivity and accuracy with immunoturbidimetry as a reference method. RESULTS The non-specific reaction was very low with the average FS less than 20 Hz, wide linear range from 0.01mg/L to 40mg/L, the limit of detection was 3μg/L, the CV of the duplicate tests and recovery tests were both less than 10% and the detection time was very short (10 min). Bland-Altman plot showed this biosensor array being highly comparable with the traditional immunoturbidimetry method. CONCLUSIONS To use biosensor array to detect the β2 microglobulin in the urine can be a feasible method in detecting trace proteins, so it can be used clinically to help diagnose early kidney injury.
What problem does this paper attempt to address?