Modification of ammonium molybdate spectrophotometric method for measurement of dissolved phosphorus

Shen Yu,Li Ling
DOI: https://doi.org/10.3969/j.issn.1002-4956.2013.01.017
2013-01-01
Abstract:The ammonium molybdate spectrophotometric method is the most commonly used for determination of phosphorus concentration in water environment.The modification of method will make it more widely application in the research field.Deeply analysis of the impact factors for determination of phosphorus concentration is made,and the time span for measuring,the optimum working wavelength and the pH range of water samples are studied.The results show that the wavelength of peak absorbance such as 710nm is regarded as the operating wavelength,and the time span for measuring is from 5 to 25 minutes at 20 ℃ temperature during the experiment.In addition,the pH value ranges from 3 to 11,which does not affect the phosphate solution,does not affect the determination of phosphorus concentration.
What problem does this paper attempt to address?