Abstract:The high-performance liquid chromatography method for detection of triphenyltin(TPhT) in sediment was optimized using samples from sediments of Lianjiang river at Guiyu town.The sediment mixed with TPhT,HCl and ethyl acetate was placed on a rotary shaker at 140r/min for 30 minutes,and further extracted by ultrasonic for 15 minutes to extract TPhT.The organic supernate was separated by centrifugation.To remove water from the supernate,anhydrous sodium sulfate was added.Then the above organic phase was evaporated nearly dry in a rotary evaporator at 40 ℃.Subsequently the residue was dissolved to constant volume with flow phase of methanol: pure water: trifluoroacetic acid as 50∶45∶5 and analyzed by HPLC to detect the concentrations of TPhT.The experiments illuminated that the calibration curve of TPhT showed a good linear relationship at the range from 0.5 to 75 mg/kg.The equation,relative coefficient of the curve and concentrations of TPhT in the sediment were y=25 354x+166 193,0.988 and(0.94±0.33) mg/kg respectively.