Longitudinal Forces Acting At Side-To-End And End-To-Side Anastomoses When A Knitted Polyester Arterial Prosthesis Is Implanted In The Dog

L Dadgar,Ar Downs,Xy Deng,Y Marois,Mw King,L Martin,R Guidoin
DOI: https://doi.org/10.3109/08941939509023139
1995-01-01
Journal of Investigative Surgery
Abstract:In a previous study, which investigated the tensions at the proximal and distal end-to-end anastomoses of a vascular prosthesis, no significant differences were found in the longitudinal forces between the two anastomotic sites after different periods of implantation. The present follow-up study was devised to study the longitudinal forces at a proximal side-to-end and a distal end-to-side anastomosis using a warp knitted polyester prosthesis implanted as a canine thoracoabdominal bypass. The external surface of the prosthesis contained a specially printed ''tension indicator'' design, which enabled changes in length to be assessed photographically at the two anastomoses at the time of implantation and at sacrifice. The longitudinal force at each sire was then calculated using a relationship obtained experimentally between the total longitudinal force and the changes in length of the virgin graft in vitro. The in vitro measurements an the prosthesis were performed using a computer-controlled laser calibration system. Although the results showed a tendency toward a gradual loss in longitudinal force at both anastomoses over 7 months in situ, statistical analysis showed no significant difference in the longitudinal force at the two anastomotic sites after any period of implantation. Histopathological and textile analysis also showed that the characteristics of the prostheses were similar at both anastomotic sites. This does not mean, however, that the intramural stress concentrations experienced by the two sites were the same. While comparing the level of the longitudinal forces measured in this study with those observed in the previous thoracic aorta study using end-to-end anastomoses, no significant differences were observed due to the different types of anastomoses. It would appear that the longitudinal force is not one of the main etiological factors that cause the formation and localization of anastomotic failure.
What problem does this paper attempt to address?