Effect of Work Parameters for Surface Ionization Source on Ion Mobility Spectrum

Zhang De-Xin,Gao Xiao-Guang,Jia Jian,He Xiu-Li,Li Jian-Ping
DOI: https://doi.org/10.3724/sp.j.1096.2011.01491
2012-01-01
Abstract:There is no reactant ion in surface ionization and sample molecules are ionized at the emitter surface. Based on this feature, an ionization source was designed in a form of a plane (the emitter) parallel to a grid (the gate). Molybdenum was used as the emitter material and triethylamine (TEA) as the test compound, the effect of the emitter temperature and the gate voltage pulse width and height on ion mobility spectrum have been investigated. The results indicate that the intensity of ion current is strongly dependant on the emitter temperature, the optimal temperature is about 400 C; the device sensitivity enhances as the gate voltage pulse width and height are increased, but the resolution lowered, the optimal pulse width and height are 100 mu s and 500 V respectively. The ionization source is easy to be fabricated with high stablility and long working life. Under the optimized parameters, the detection limit of triethylamine is less than 0. 04 mg/m(3),
What problem does this paper attempt to address?