A Fast Solution Switching System with Temperature Control for Single Cell Measurements

Duk-Su Koh,Liangyi Chen,Carmen A. Ufret-Vincenty,Seung-Ryoung Jung
DOI: https://doi.org/10.1016/j.jneumeth.2011.04.023
IF: 2.987
2011-01-01
Journal of Neuroscience Methods
Abstract:This article describes a perfusion system for biophysical single cell experiments at the physiological temperature. Our system regulates temperature of test solutions using a small heat exchanger that includes several capillaries. Water circulating inside the heat exchanger warms or cools test solutions flowing inside the capillaries. Temperature-controlled solutions are delivered directly to a single cell(s) through a multibarreled manifold that switches solutions bathing a cell in less than 1s. This solution exchange is optimal for patch clamp, single-cell microamperometry, and microfluorometry experiments. Using this system, we demonstrate that exocytosis from pancreatic β cells and activation of TRPV1 channels are temperature sensitive. We also discuss how to measure local temperature near a single cell under investigation.
What problem does this paper attempt to address?