Effect of different activating buffer systems on determination of alkaline phosphatase activity

ZHENG Tiesheng,YE Lixin,WU Zhiyong,ZHANG Xu,SHENG Weixiang
DOI: https://doi.org/10.3969/j.issn.1001-764X.2005.06.008
2005-01-01
Abstract:Objective To investigate the effect of different activating buffer systems on the activity of alkaline phosphatase,and to provide reference for clinical diagnosis.Method We determined alkaline phosphatase activity in serum samples collected from clinical patients and normal donors using three methods based on different activating buffer systems.Results The three methods in which DAE,EAE or AMP was used as activating buffer systems appeared to have normal reference range of 98±57,86±44 and 75±42 U/L respectively.The comparison among the results of three methods showed significant difference(P<0.05).The determined values of alkaline phosphatase activity of seven groups with liver disease including 221 patients using the three activating buffer systems were significantly higher than those of normal persons(P<0.001).The values of alkaline phosphatase activity between the patients and normal persons determined by the three methods were significantly different(P<0.05).Conclusion Different activating buffer systems in the determination obviously influence the results of serum alkaline phosphatase activity.Therefore,selection of optimal activating buffer system is very important for measuring alkaline phosphatase activity although the substrate is same.It is worthy to notice that the methods using different activating buffer systems may have different normal reference range.
What problem does this paper attempt to address?