Determination of Pre-column Dansylation of 4-methylimidazole by HPLC

Wang Miaofei,Wang Yonghua,Guo Xindong,Du Zhifeng,Wu Yuluan,Zhang Shuihua
2007-01-01
Abstract:A method for determination 4-methylimidazole by using high-performance liquid chromatographic (HPLC)was developed.This method was based on pre-column derivatization with dansyl chloride(Dns-Cl).The de- rivatives were separated on a C_(18) column,and eluted with acetonitrile and 0.02 mol/L ammonium acetate solution by a gradient eluate,monitored by fluorescence detection.The method was fast and accurate with a linear correlation of 0.998 5 in the range of 0.05 to 1.0 mg/L,average recoveries of 71.6% to 88.7%,a relative standard deviation of 3.6 % to 5.5 % and a minimum determination concentration was 10.0μg/L.
What problem does this paper attempt to address?