High-resolution OCT Balloon Imaging Catheter with Astigmatism Correction.

Jiefeng Xi,Li Huo,Yicong Wu,Michael J. Cobb,Joo Ha Hwang,Xingde Li
DOI: https://doi.org/10.1364/ol.34.001943
IF: 3.6
2009-01-01
Optics Letters
Abstract:We report new optics designs for an optical coherence tomography (OCT) balloon imaging catheter to achieve diffraction-limited high resolution at a large working distance and enable the correction of severe astigmatism in the catheter. The designs employed a 1 mm diameter gradient-index lens of a properly chosen pitch number and a glass rod spacer to fully utilize the available NA of the miniature optics. Astigmatism caused by the balloon tubing was analyzed, and a method based on a cylindrical reflector was proposed and demonstrated to compensate the astigmatism. A catheter based on the new designs was successfully developed with a measured diffraction-limited lateral resolution of approximately 21 microm, a working distance of approximately 11-12 mm, and a round-shape beam profile. The performance of the OCT balloon catheter was demonstrated by 3D full-circumferential imaging of a swine esophagus in vivo along with a high-speed, Fourier-domain, mode-locked swept-source OCT system.
What problem does this paper attempt to address?