An Ni(Chitin)(2) Modified Nitric Oxide Microsensor

Yuezhong Xian,Jian Xue,Song Zhang,Xiangyang Ying,Litong Jin
DOI: https://doi.org/10.1007/s002160051527
1999-01-01
Abstract:The development of a simple, sensitive, selective, and stable amperometric nitric oxide microsensor is described. It is based on Ni(chitin)2 mediators immobilized on a platinum, Nafion modified electrode. The detection of NO is based on the Ni(chitin)2 catalysis of NO oxidation at an anodic potential of +0.74 V (vs. SCE). The catalytic peak current is linear for a NO concentration in the range of 8.5 × 10–8 mol/L to 1.5 × 10–5 mol/L, with a correlation coefficient of 0.9992. The detection limit of the microsensor is 5.0 × 10–8 mol/L. It is suitable for the direct measurement of NO in biological systems.
What problem does this paper attempt to address?