Quantitative Comparison Between Microfluidic and Microtiter Plate Formats for Cell-Based Assays

Huabing Yin,Nicola Pattrick,Xunli Zhang,Norbert Klauke,Hayley C. Cordingley,Steven J. Haswell,Jonathan M. Cooper
DOI: https://doi.org/10.1021/ac701958z
IF: 7.4
2007-01-01
Analytical Chemistry
Abstract:In this paper, we compare a quantitative cell-based assay measuring the intracellular Ca2+ response to the agonist uridine 5'-triphosphate in Chinese hamster ovary cells, in both microfluidic and microtiter formats. The study demonstrates that, under appropriate hydrodynamic conditions, there is an excellent agreement between traditional well-plate assays and those obtained on-chip for both suspended immobilized cells and cultured adherent cells. We also demonstrate that the on-chip assay, using adherent cells, provides the possibility of faster screening protocols with the potential for resolving subcelluar information about local Ca2+ flux.
What problem does this paper attempt to address?