Detection of pepsinogenⅡby three different methods and its clinical application

Zhao Fang,Zhao Zhen,Pan Huifen
DOI: https://doi.org/10.3969/j.issn.1673-4130.2013.19.010
2013-01-01
Abstract:Objective To evaluate the correlation of three pepsinogenⅡ(PGⅡ)detection methods and its clinical applications.Methods 88serum samples were randomly selected,and PGⅡwas detected by using time-resolved fluoroimmunoassay(TRFIA),chemiluminescence immunoassay(CLIA)and enzyme-linked immunosorbent assay(ELISA),respectively.Then,their results were processed using regression and correlation analysis.To verify the feasibility of gastric disease screening by PGⅡdetection,5 439 people were divided into three groups:gastroenterology outpatients,gastroenterology hospitalized patients,and physical examination population.Results It showed a significant positive correlation between any two of the three methods(P<0.05).PGⅡserum levels were(15.97±10.82)ng/mL,(20.75±17.27)ng/mL,and(35.64±41.14)ng/mL in physical examination population,outpatients,and hospitalized patients,respectively.There was significantly difference between the three groups(P<0.05).Conclusion PGⅡdetection may be used for the screening of superficial gastritis and gastroduodenal ulcer.But the laboratory should establish its own reference values according to different methods.
What problem does this paper attempt to address?