Carotid artery stenting with cerebral protection device for carotid stenosis

BAO Jun-min,MEI Zhi-jun,LU Qing-sheng,FENG Xiang,ZHAO Zhi-qing,JING Zai-ping
DOI: https://doi.org/10.3969/j.issn.1674-7429.2009.02.007
2009-01-01
Abstract:Objective To summarize our experiences in carotid artery stenting(CAS) with cerebral protection device for stenosis and to study its periprocedural management.Methods One hundred and twenty-five CAS procedures were performed on 122 patients(99 men;mean age 72.1 years,range 52-87) from January 2002 to December 2007 and the data were retrospectively reviewed.Results One hundred and twenty-four stents were successfully deployed in 123 arteries(technical success rate 98.4%).Various cerebral protection devices were used in all cases.In the periprocedural period,6 transient ischemic attacks(TIAs)(4.9%),2 minor strokes(1.6%),1 strokes(0.8%),2 severe reperfusion injuries(1.6%) and 0 death(0%) occurred.The all stroke/neurological death rate was 2.4%.No neurological events were observed in 86 cases during a mean follow-up period of 18.5 months(range 3-36),although 5 patients with 50%~70% restenosis(5.8%) and one with >70% restenosis(1.2%)were detected.Two patients died from unrelated causes.The other 78 cases(90.7%) were not observered stent restenosis and neurological complications.Conclusions CAS with cerebral protection device is a safe and effective method for carotid stenosis.Improvement of periprocedural management is critical to improve outcomes and reduce complications.
What problem does this paper attempt to address?