Biomechanical properties of anterior column plate combined with posterior column lag screw in fixing complicated acetabular fractures
Lie-hu CAO,Qi-rong ZHOU,Wei-zong WENG,Di LI,Qiang WEI,Xiao CHEN,Jin CUI,Jun ZHANG,Lin WANG,Fang JI,Jia-can SU
DOI: https://doi.org/10.16781/j.0258-879x.2017.11.1397
2017-01-01
Abstract:Objective To compare the mechanical stability and efficacy of anterior column plate combined with posterior column lag screw and double colunm plate in fixing the complicated acetabular fracture.Methods Six pelvic specimens retaining ligaments and hip joint capsules from adult males were selected to construct the acetabular T fracture models,with three fracture lines at the anterior column,posterior column and quadrilateral area.The acetabular fracture model was fixed with anterior column plate combined with posterior colunn lag screw on the right side (experimental group),and was fixed with double column plate on the left side (control group).A total of six pairs of marking points,with two pairs each in the anterior column,posterior column and quadrilateral aera,were recorded as 1 and 2,3 and 4,5 and 6 according to the order of anterior column,posterior column and quadrilateral area on both sides.The specimens were fixed on a biomechanical pressurizer for axial compression,with a vertical loading pressure from 0 N to 1 500 N,circulating 6 times;the vertical displacement (difference of the distance between the two points of the pair of marking points before and after the load) of each pair of marking points were recorded at 300 N,600 N,900 N,1 200 N,and 1 500 N.The average vertical displacement and shear stiffness of the specimens were calculated and compared in the two groups.Results No fracture or internal fixation fracture was observed in the 6 cycles axial loading.The longitudinal displacement of each pair of marking points in the two groups were increased with the increase of load,but the maximum longitudinal displacement did not exceed 1 rnm,and there was no significant difference between the two groups.The shear stiffness of the specimens in the experimental group was less than that in the control group ([1 428±-57] N/mm vs [1 621±62] N/mm),but there was no significant difference between the two groups (P>0.05).Conclusion Anterior column plate combined with posterior column lag screw (with the load of 300 N,600 N,900 N,1 200 N,1 500 N) can effectively fix T fracture and meet the requirements of fracture fixation,which is stable and reliable,and is similar to conventional double column plate in fixing the complicated acetabular fracture.