A Preliminary Study of Metal Fixation Artifacts by Spectral CT

BI Zheng-hong,WANG Ming-Peng,MAO Ding-biao,BAI Ai-guo,NI Zhi-hua,HU Ya-min,XU Jie,LI Ke
DOI: https://doi.org/10.3969/j.issn.1006-5741.2013.01.018
2013-01-01
Abstract:Purpose:To explore the image quality of fractures after metal fixation by spectral CT.Methods:Forty-five patients with metal implants were scanned with GSI(Gemstone spectral CT) examination.The spectrum analysis was used to select the monochromatic images at 120 keV+MARS.After the postprocessing was done,the image quality of MPR,VR,MIP was evaluated.Results:The image quality of MPR,VR and MIP met the diagnostic requirement.Image quality was excellent in 86.3% of the patients,and 84.0% of them was with nearly no image artifacts.Conclusion:The image quality of fractures after metal fixation can be obviously improved and the fine structure can be clearly displayed by spectral CT through the remove of metal artifacts and beam-hardening artifacts.