Optimization of Measuring Apparatus for Pressure-composition-temperature Relationship in Hydrogen Storage Alloys

Tongjin Chen,Zhu Wu,Pei Qin,Zhilin Li,Xuebin Yu,Jianfeng Mao,Tiesheng Huang,Neixin Xu
DOI: https://doi.org/10.3969/j.issn.1000-0380.2008.02.020
2008-01-01
Abstract:As the hydrogen storage alloys have been used in wider energy field,pressure-composition-temperature(PCT) measurement apparatus is required to be more accurate and more reliable.Aiming at this target,the optimized strategy for PCT apparatus is proposed.The accuracy of data acquisition and stability of operation have been enhanced by adopting 2.5V external reference voltage,and ADμC848 controller,in which 16-bit resolution ADC is integrated.The measuring principle based on gas state equation as well as the structure of device are described briefly.In addition,the performance of controller,K-type thermocouple circuit,hardware configuration and software flowchart are expounded in detail.The experimental operation of this optimized apparatus indicates that reliable test of full performance of hydrogen storage alloys can be implemented with this device.