Design and implementation of boron measurement instrument using time-resolved fluorescence technology based on C8051F020

Chu Zhulin,Shi Chun,Zhou Quan,Wu Gang
DOI: https://doi.org/10.3969/j.issn.1002-7300.2010.09.024
2010-01-01
Abstract:Boron measurement instrument using Time-resolved fluorescence technology based on C8051F020 is developed.C8051F020 is used as core controller in the system;the output PWM signal is used to control pulse light excitation issued;ADC conversion module is used to acquisition fluorescence signal of boron in the material;the output DA signal is adjusted to control negative high voltage power supply of photomultiplier according to the data acquisition,so photomultiplier's output current can be kept in a reasonable extent to improve the accuracy of data acquisition.Boron measurement instrument communicates with PC via the RS232 bus,and sends the data of fluorescence of boron to PC for further processing.keyboard and LCD are used as friendly interface.Experimental results show that,due to using a modular design,Boron measurement instrument has advantages of structural flexibility,scalability and cost-effectiveness.So it has a wide range of practicality.
What problem does this paper attempt to address?