[Accuracy of Three-Dimensional Facial Image Based on Phase Measuring Profilometry].

Yi Liu,Li-li Ma,Si Chen,Tian-min Xu
DOI: https://doi.org/10.3760/j.issn:1002-0098.2007.01.012
2007-01-01
Abstract:OBJECTIVEThree-dimensional (3-D) facial images are very useful for orthodontic diagnosis. Three-dimensional facial imaging system (DSC-2) based on phase measuring profilometry (PMP) was used to obtain images and tested for accuracy.METHODSGeometric strips (20 mm in width) was used to test the accuracy of the obtained images. Three-dimensional images of plaster head model with marker of 5 mm x 5 mm were taken by DSC-2 three-dimensional facial imaging system. The difference of measurements was compared between both methods. Measurements were repeated twice by three different doctors.RESULTSThe average value of width in 3-D mode was 19.5 mm. No significant difference was found, compared with actual width (20 mm). There was no significant difference between computer measurement and manual measurement of the plaster head model (P > 0.05).CONCLUSIONSDSC-2 three-dimensional facial imaging system was accurate and reliable to obtain 3-D facial images.
What problem does this paper attempt to address?