Evaluation of the Reliability of Thickness Measurement of Mouse Brain Frozen Sections with Stereo Investigator 8.0 System

ZHOU Yan,WANG Dongxin,YU Jianchun,CHENG Haiyan,YUAN An
2012-01-01
Abstract:Objective To evaluate the reliability of thickness measurement of mouse brain frozen sections with Stereo Investigator8.0 system.Methods Four-month-old(weighing 20-30 g) male C57BL/6J mice were randomly divided into two groups: 40 μm and 60 μm groups.The mice were anesthetized and fixed with paraformaldehyde perfusion,and the brains were taken for frozen sectioning.Stereo Investigator 8.0 system was used to measure the thickness of each selected sections at a fixed spot repeatedly by one person or by different persons.Results The intraclass correlation coefficients of the 40 μm and 60 μm groups carried out by one person were 0.897 and 0.988,respectively,and the coefficient of the 60 μm group carried out by three persons was 0.895.However,in the 40 μm group carried out by three persons,the intraclass correlation coefficient was only 0.658.Conclusions The thickness measurement of frozen sections with Stereo Investigator 8.0 system shows a good reliability in 40 μm and 60 μm groups carried out by one person,and 60 μm group carried out by three persons.However,the measurement reliability seems questionable in the 40 μm group carried out by three persons.