Modification of Miniature Transducer Element for Forward Beam Scanning

HU Bing,ZHU Jia-an,SHOU Wen-de,QIAN De-chu,FENG Liang,ZHU Ling
DOI: https://doi.org/10.3969/j.issn.1000-3630.2005.04.011
2005-01-01
Abstract:A standard miniature transducer element is modified and its practicability for forward scanning evaluated.The standard element that launches a beam at 90° with respect to the axis of the catheter has been modified to deviate the beam in the forward direction at 30° or 45° to the axis Experiments have been carried out to validate the desired modification.The deviated scanning beam was successfully measured,and the primary acoustic parameters such as impedance and direction were slightly changed.The modified transducer can reach a range extended up to 2.893.46mm,overrunning the length of metal cecum.The modified miniature transducer can detect lesion in the forward direction of the catheter.With this modification,the problem of ultrasonic detection when stricture occurs and neoplasms are located in dead zone of the upper renal calices can be solved.