Refractive index dispersion measurement on nonlinear optical polymer using V-prism refractometer

Wei Shi,Changshui Fang,Xin Yin,Qiwei Pan,Xun Sun,Qintian Gu,Jinzhong Yu
DOI: https://doi.org/10.1016/S0143-8166(99)00047-0
IF: 5.666
1999-01-01
Optics and Lasers in Engineering
Abstract:By using V-prism refractometer, the refractive indices of a polyetherketone (PEK-c) guest–host polymer system were measured with the polymer in solutions. The Lorenz–Lorentz local field formalism was used in the calculation of the refractive indices of the polymers from the measured indices of the polymer solutions and the pure solvent by using V-prism refractometer. The refractive index dispersions of the polymers were obtained by fitting the measured indices of the polymers to Sellmeyer equation. The method allows for an accuracy in index of 0.7% in the determination of the polymer indices. In addition, a large difference between the indices of the polymer and the solvent, and a higher polymer volume fraction in the measured polymer solution are favorable for a high accuracy.
What problem does this paper attempt to address?