Determination of Bovine Serum Albumin by Resonance Light Scattering Technique

牧丹,吴立航,高德江,孙滟涛,邓新煜,田媛,张寒琦
2008-01-01
Abstract:In this paper,the bovine serum albumin(BSA)-β-mercaptoethanol(β-ME)-sodium dodecylbenzene sulphonate(SDBS) system was studied by resonance light scattering technique.Based on the RLS intensity difference between β-ME-SDBS system and BSA-β-ME-SDBS system,a simple assay for BSA was developed.The influences of some experimental factors,including pH value,concentration of β-ME and SDBS,addition sequence of reagents and foreign substances,on the enhancement of the RLS intensity were studied.Under the optimal conditions,the enhanced RLS intensity was proportional to the concentration of BSA in the range from 1.0×10-8 mol\5L-1 to 7.0×10-7 mol\5L-1 with the detection limit of 6.0×10-9 mol\5L-1.Some metal ions had little effect on the determination of BSA.The results of assay for BSA in synthetic samples and real sample were satisfactory,which were in good agreement with the results obtained by Bradford method.
What problem does this paper attempt to address?