Feasibility Studies on Low-dose Cone-beam 3-D Mammography

Zhengrong Liang,Hongbing Lu,Junhai Wen,Tianfang Li,Wei Zhao,Jiang Hsieh,Ge Wang,Roxanne Palermo
2003-01-01
Abstract:Breast cancer is the most frequent cause of deaths for women in USA. Currently available digital mammography provides excellent spatial resolution and tissue contrast with a reasonable soft X-ray radiation dosage. With further technology development, the resolution and contrast would be further improved toward early detection of the cancer. A main limitation of the conventional digital mammography is the overlap of tissues along the projection of X-rays onto the detector, in addition to the complication for intervention because of the structural distortion associated with the breast compression. Three-dimensional (3D) mammography has been investigated in past years, aiming to remove the overlap limitation. A significant problem for 3D mammography is its increased radiation, as compared to the conventional mammography. This work aims to reduce the radiation, while achieving high quality 3D mammography, by investigating (1) an optimal system configuration for as complete as possible data sampling with least projections and (2) an effective noise reduction on the acquired data with as less as possible X-rays. An optimal data sampling orbit was suggested after investigating several candidates with cone-beam geometry. An effective noise reduction framework was established, which could achieve more than ten times radiation reduction while retaining the same image quality. Further investigation and validation are under progress.