CH4 Gas Remote Sensing by a Transceiver System Based on TDLAS Technology

姜治深,王飞,许婷,邢大伟,姚华,严建华,岑可法
DOI: https://doi.org/10.3969/j.issn.1004-3950.2012.03.001
2012-01-01
Energy Engineering
Abstract:A telescope transceiver system design principle based on TDLAS technology was described.Through the measurements of CH4 concentration in the absorption cell,the accuracy of the measurement was calibrated.On this basis,by detecting CH4 concentration of different work conditions in a 40m-long open path,it was verified that the system had a good time responsiveness.The minimum detectable limit of the system was about 5.48×10-5 m/m.Studies showed that the system could meet the demand of on-line detection of industrial gases.In addition,the system could be also used to detect the other hazardous gas concentration,such as H2S,CO,and NH3.