Accuracy assessment of computer aided stimulation systerm for orthognathic surgery

SONG En,MO Kangnan,ZHU Fudong,GE Weili,LI Shenglai,XIE Zhijian
DOI: https://doi.org/10.13591/j.cnki.kqyx.2013.01.010
2013-01-01
Abstract:Objective To assess the accuracy of computer aided stimulation systerm(SASSOS 2001) for orthognathic surgery.Methods The study consisted of 25 patients with maxillofacial deformity treated with orthognathic surgery.Their lateral cephalograms and color lateral photos of soft tissue before surgery were input to the CASSOS 2001 system.Carry out cephalometric analysis and simulative orthognathic surgery by bone movement and predict occlusal relationship and lateral features of soft tissue.Then input the lateral cephalograms after surgery into the CASSOS 2001 system,and carry out cephalometric analysis.Choose ten parameters and compare predicted surgical parameters with actual postsurgical parameters.Results Results of four parameters of hard tissue and five parameters of soft tissue in this study were similar(P>0.05).There were statistical differences in only one parameter of soft tissue,the labral length(P<0.05).Conclusions CASSOS 2001 system is precise for the prediction simulation of changes of hard tissue in orthognathic surgery.It has a little shortcoming in the prediction of soft tissue,especially in the prediction of lips.The positioning error of the marking points,and the soft to hard tissue movement proportion have a large influence in the prediction.CASSOS 2001 system still has great room for improvement.
What problem does this paper attempt to address?