Determination of vanillin with potassium permanganate-luminol chemiluminescence system

ZHOU Yan-mei,WANG Ya-ping,MA Tong-sen,XIE Xiao-long
DOI: https://doi.org/10.13624/j.cnki.issn.1001-7445.2012.03.012
2012-01-01
Abstract:Based on the fact that the chemiluminescence(CL) reaction of potassium permanganate and luminol in alkaline medium could be enhanced by vanillin,a new analytical method of chemiluminescence to detect vanillin was developed,which expands the application of flow-injection chemiluminescence in food analysis.The effect of type of medium and concentration of reagents on CL intensity was investigated.The results showed that the optimum concentrations of NaOH,KMnO4 and Lumino were 0.20 mol/L,3.0 × 10-5 mol/L and 7.0 × 10-5 mol/L,respectively.Under the optimum conditions,linearity relationship of values of CL intensity and mass concentration of vanillin was obtained in the range of 5.0 × 10-8~3.0 × 10-6 g/mL,with a equation of linear regression as ΔI = 7.188 3 × 108 C(g/mL) + 71.767 4(r = 0.994 2).Parallel determination of 5.0×10-7 g/mL vanillin for 11 times show that the RSD is 1.3% with a detection limit of 1.1×10-8 g/mL.A new method of flow-injection chemiluminescence to detect vanillin was found,which has been successfully applied to the determination of vanillin in pill and the recovery rate obtained were in the range of 98%~104%.
What problem does this paper attempt to address?