Electrochemical Determination of Ascorbic Acid in Room Temperature Ionic Liquid BPPF6 Modified Carbon Paste Electrode

Wei Sun,Maoxia Yang,Ruifang Gao,Kui Jiao
DOI: https://doi.org/10.1002/elan.200703889
IF: 3.077
2007-01-01
Electroanalysis
Abstract:A room temperature ionic liquid N-butylpyridinium hexafluorophosphate (BPPF6) was used as a binder to make an ionic liquid modified carbon paste electrode (IL-CPE), which showed good characteristics such as simple preparation procedure, fast electrochemical response and good conductivity. ne electrochemical oxidation of ascorbic acid (AA) on the new IL-CPE was carefully studied. The oxidation peak potential of AA on the IL-CPE appeared at 109 mV (vs. SCE), which was about 338 mV decrease of the overpotential compared to that obtained on the traditional carbon paste electrode (CPE) and the oxidation peak current was increased for about four times. The electrochemical parameters of AA on the IL-CPE were calculated with the charge transfer coefficient (alpha) and the electrode reaction rate constant (k(s)) as 0.87 and 0.800 s(-1), respectively. Based on the relationship of the oxidation peak current and the concentration of AA a sensitive analytical method was established with cyclic voltammetry. ne linear range for AA determination was in the range from 1.0 x 10(-5) to 3.0 x 10(-3) mol/L with the linear regression equation as I-P (mu A) = -2.52 - 0.064C (mu mol/L) (n = 13, gamma = 0.9942) and the detection limit was calculated as 8.0 x 10(-1) mol/L (3 sigma). The proposed method was free of the interferences of coexisting substances such as dopamine (DA) and amino acids etc., and successfully applied to the vitamin C tablets determination.
What problem does this paper attempt to address?