Retrospective analysis on reexamination of infective markers of pretransfusion testing by chemiluminescent microparticle immunoassay
ZHANG Min,LIU Feng,LI Guining,LI Minfang
DOI: https://doi.org/10.13201/j.issn.1004-2806-b.2013.01.010
2013-01-01
Abstract:Objective:To assess the reproducibility of detecting infective markers of pretransfusion testing by chemiluminescent microparticle immunoassay(CMIA),and develop a reasonable strategy of reexamination. Method:Blood samples were reexamined due to low concentration of HBsAg or rare models in combination of HBV markers from May 2010 to Nov 2011 in our hospital by CMIA including 436 cases of HBsAg,194 cases of HBsAb,147 cases of HBeAg,161 cases of HBeAb,85 cases of HBcAb,756 cases of HCV,707 cases of Syphilis and 94 cases of HIV,and their concordance were assessed. Result:The coincident rate in marker of HBsAg was 63.07%.The coincident rates of HBsAg whose concentration≤0.1 U/mL,0.1-1.0 U/mL,1-10 U/mL and >10 U/mL were 38.25%,66.07%,84.21% and 96.11%,respectively,and the significant differences were found in all the above groups(P<0.01).By further grouping,the coincident rates of HBsAg whose concentration between 0.05-0.15 U/mL and 0.15-1.00 U/mL had significant differences from statistics(P<0.05).The coincident rates of HBsAb,HBeAg,HBeAb,HBcAb,HCV,Syphilis and HIV were 96.39%,91.16%,96.27%,96.47%,97.35%,98.73% and 91.49%,respectively.The coincident rates of HCV whose values of S/CO between 0.90-1.09 and 1.10-1.59 had significant differences(P<0.01). Conclusion:The reproducibilities were good in detection of HBsAb,HBeAg,HBeAb,HBcAb,HCV,Syphilis and HIV by CMIA,rather than in the detection of HBsAg.When HCV whose values of S/CO were 0.90-1.09,or HBsAg whose concentration less than 0.15 U/mL,the reexaminations should be performed with high speed centrifugation to ensure the accuracy.