Evaluating the application of flow fluorescence assay for neuron-specific enolase detection

WU Chuan-yong,XU Shu-jun,XUE Jian,GE Xin-yue,ZHU Xiao,WU Jing,LOU Jia-tao
2012-01-01
Abstract:To establish normal reference value for neuron-specific enolase(NSE) by flow fluorescence assay and evaluate the potential application to clinical use,the levels of NSE were detected by flow fluorescence assay and electrochemiluminescence immunoassay(ECLIA) using a commercial detenction kit,respectively in 176 clinical serum samples to determine correlation between the two methods.The serum NSE levels of 668 healthy controls were detected by flow fluorescence assay in order to establishing normal reference values for NSE.And then,NSE levels of 242 cases of lung cancer,108 benign lung diseases and 147 healthy controls were detected by flow fluorescence assay for evaluating the application of flow fluorescence assay for NSE detection.The NSE values detected by these two methods for 176 clinical serum samples have fine correlation(Y=0.9746X-0.2075,R2=0.9653).The normal reference value for NSE by flow fluorescence assay was determined according to the first 95% percentile of the values from 668 healthy controls(17.07 ng/ml),and the reference value was comparable to that by ECLIA.The NSE positive rate(>17.07 ng/ml) in small cell lung cancer cases was significantly higher than that in lung squamous cell carcinoma cases and lung adenocarcinoma cases(P<0.01).In conclusion,the normal reference value for NSE by flow fluorescence assay has beens established,and its application to NSE detection has been preliminarily evaluated,which lay the foundation for routine application in NSE detection.