Comparison of Performance of Matrix-Assisted Laser Desorption/Ionization-Time of Flight-Time of Flight and Electrospray Ionization-Quadrupole-Time of Flight for Qualification and Quantification of Proteins

Ni Guo-Xin,Zhu Lei,Ma Xiao-Tu,Xu Xue-Min,Hu Xiao-Fang,Liu Jian-Hua,Li Wei
DOI: https://doi.org/10.3321/j.issn:0253-3820.2009.09.008
2009-01-01
Abstract:To compare the performance of MALDI-TOF-TOF and ESI-Q-TOF in protein qualification and quantification, isobaric tags for relative and absolute quantification (iTRAQ) labeled peptides derived from trypsin-digested proteins in MCF7 cells with or without estrogen treatment were identified and quantified using MALDI-TOF-TOF and ESI-Q-TOF respectively. The experimental result showed that 1086 and 848 proteins were identified by MALDI-TOF-TOF and ESI-Q-TOF respectively, wherein 633 proteins were homolog; 38 and 33 proteins were found to be differentially expressed by more than 0.5 fold by MALDI-TOF-TOF and ESI-Q-TOF respectively, wherein 3 proteins were the same proteins. The experimental data indicate that these two kinds of instruments were complementary both in peptide identification and relative quantification. Paralleled use of MALDI-TOF-TOF and ESI-Q-TOF significantly increased both the coverage and the confidence level of identified and quantified proteins.