Variety of Respiratory Parameters under Pediatric Tidal Volume in Some DatexOhmeda Anesthesia Machines

YANG Yun,WANG Xiao-hai
2012-01-01
Abstract:Objective To discuss the variety of the common respiratory parameters under the control mode of pediatric tidal volume of some DatexOhmeda anesthesia machines.Methods Eleven DatexOhmeda anesthesia machines were used for clinical simulated study with the desired tidal volume of 100mL,200mL or 300mL.The oxygen flow was set as 2L/min,with corrugated tubes and breathing bags of the children and adults mobilized.Such parameters were measured as the number of bellows(VT),actual tidal volume(VTE),minute ventilation(VE),indication of airway pressure table(pmax1),indication of the display(pmax2) and mean airway pressure(pmean).Results When the corrugated tube of the children mobilized,there were a VT value higher than desired one,a VTE value less than VT;the VT value and VE value were all higher than those when the corrugated tube of the adults mobilized.Aestiva/5(7900) machine had a VT value higher than the desired one when using pediatric corrugated tube.Aestiva/5(7900) machine had higher VT values with the preset tidal volumes of 200mL and 300mL,and Aestiva(7100) machine had higher VT values when mobilizing pediatric corrugated tube than adult one.Aestiva/5(7900) machine had higher VTE values with the preset tidal volumes of 100mL and 300mL,and Aestiva(7100) machine had higher VTE values when mobilizing pediatric corrugated tube than adult one.Conclusion The accuracy of the machine is satisfactory,pediatric corrugated tube and breathing bag are necessary for pediatric anesthesia,and Anestiva/5(7900) is fit for pediatric anesthesia.