Systematic Evaluation of Hardware and Animal Stability for High-Resolution Layer-Specific MRI of the Retina

Timothy Q. Duong,Xiaodong Zhang,Yingxia Li
2009-01-01
Abstract:Purpose: To examine animal and hardware instability in phantoms and in vivo rat retinas, and to implement solu- tions to improve stability for structural and functional MRI of the retina. Materials and Methods: Anatomical (25 25 m) and perfusion-based functional (90 90 m) MRI protocols were evaluated on phantoms and in vivo rat retinas at 7 Tesla. Temporal phase drift and pixel displacement were evaluated. Comparisons were made before and after phase correction and spatial co-registration. Results: The retina drifted slightly over time despite under deep isoflurane anesthesia. Combined paralytic and 1.1%- isoflurane eliminated eye movement, as confirmed by opti- cal imaging. To maximize hardware stability, the animal holder was fastened to the gradient, and high-order shims were not used. Nonetheless, small but appreciable image displacement along the phase-encode, but not the readout, direction was observed. Such displacement was likely due to Bo-drift and was removed by postprocessing phase cor- rection. By comparison, brain MRI at the typical resolution (400 400m) showed no significant displacement, likely because lower resolution has less gradient demand and displacement is less apparent. Conclusion: We concluded that optimizing animal and hardware stability is important for high-resolution retinal MRI. MRI could readily provide a valuable tool to study animal models of retinal diseases.