Characterization of impurities in semi-synthetic vinorelbine bitartrate by HPLC-MS with mass spectrometric shift technique.

Xiaoji Cao,Yuanpo Tai,Cuirong Sun,Kuiwu Wang,Yuanjiang Pan
DOI: https://doi.org/10.1016/j.jpba.2005.02.042
IF: 3.571
2005-01-01
Journal of Pharmaceutical and Biomedical Analysis
Abstract:A simple and sensitive method of high-performance liquid chromatography coupled with electrospray ionization mass spectrometry (HPLC/ESI-MS) was developed to separate and identify impurities in semi-synthetic vinorelbine bitartrate sample. The analytical HPLC was carried out on a reversed-phase C8 column using 0.02M ammonium formicate buffer (pH 4.2) and methanol (46:54, v/v) as mobile phase at a flow rate of 0.8ml/min at room temperature and a UV detection at 267nm. The on-line HPLC/ESI-MS/MS was performed using ion trap analyzer in positive ion mode. Applying mass spectrometric shift technique to HPLC/ESI-MS/MS analysis, four impurities were identified as 18′-O-demethylvinorelbine (impurity-1), 6′-N-methylvinorelbine (impurity-2), 23-O-demethylvinoreline (impurity-3) and 17-bromovinorelbine (impurity-4), respectively, in investigated vinorelbine bitartrate sample. The four impurities, in which the impurity-1 was not reported as the semi-synthetic process impurity for vinorelbine bitartrate elsewhere, were isolated by preparative high-performance liquid chromatography. Their structures were further confirmed by means of 1D and 2D NMR spectra. Structural elucidation by spectral data was discussed.
What problem does this paper attempt to address?