The Design of Automatic Gripper Used to Clamp Micro Stitch

LI Jian-ping,YANG Yang,XU Cun-liang
DOI: https://doi.org/10.3969/j.issn.1006-2343.2012.01.030
2012-01-01
Abstract:According to the characteristics of corneal suturing operation and the technical requirements of automatic gripper used to clamp the micro stitch,the structure of an automatic gripper was designed in this paper.Based on the Hertz theory,the contact problem between two cylinders was discussed.And depending on the finite element analysis of the gripper model,the limit displacement of the linear motion joint of the gripper was obtained without plastic deformation of the needle,which providing the basis for grippers control.Finally,the operation experiment of clamping micro stitch is carried out.Results from experiments testify the feasibility and practicability of the presented automatic gripper.
What problem does this paper attempt to address?