Effects of Tetramethylpyrazine-Eluting Stent on Neointimal Hyperplasia after Renal Artery Stenting

凡六章,朱伯谦,陈龙,丁震,马根山,鞠成伟,朱建,李拥军,汤成春,陈忠
DOI: https://doi.org/10.19460/j.cnki.0253-3685.2012.14.004
2012-01-01
Abstract:Objective To evaluate the short-term effects of tetramethylpyrazine-eluting stent(TES) in preventing neointimal hyperplasia after renal artery stenting.Methods Ten healthy miniswines were equally randomized into two groups of TES(baring TES in unilateral renal artery) and BMS(baring metal stent as the control).Quantity renal arteriography(QRA) and intravascular ultrasound(IVUS) were performed before,at 15 minutes and 35 days after operation,and then the swines were sacrificed and the neointimal hyperplasia was measured by histomorphology assessment.Antiproliferating cell nuclear antigen(PCNA) antibody was used to observe smooth muscle cell(SMC) proliferation in the neointima.The coverage of endothelium was observed under scanning electron microscope(SEM).Results The late lumen loss in group TES was(0.32±0.11) mm,which was similar to(0.50±0.19) mm in group BMS(P0.05).There was no significant difference in the average percentage of area stenosis between two groups [(9.0±2.6)% vs.(11.7±2.5)%].Histomorphometry of the neointimal area changes and PCNA in two groups were similar(P0.05).SEM examination showed satisfactory endothelial coverage in both groups.Conclusion The TES has a tendency of inhibiting the neointimal hyperplasia without affecting healing up of endothelium after renal artery stenting.