Optimal Design of Prosthetic Alignment Angle for Monolithic Prosthetics

JIA Xiaohong,LI Xiaobing,WANG Rencheng,JIN Dewen
DOI: https://doi.org/10.3321/j.issn:1000-0054.2007.05.009
2007-01-01
Abstract:The two alignment angles between the prosthetic socket and the shank must be specified when a monolithic prosthetics is designed.This paper describes an optimal design model using alignment angles in the sagittal plane and the frontal plane as optimization variables with the optimal objective function to minimize the deviation of three biomechanical parameters of the lower limb related to misalignment effects.Gait analysis experiments and a three-dimensional rigid-body dynamics model for a trans-tibial amputee define two optimal alignment angles,0.32° for the knee flexion and 2.85° for the socket adduction using double interpolation.The results match well with the experimental results.This optimization algorithm can be used for numerical designs of monolithic prosthetics and for setting alignment criteria for traditional modular prostheses.
What problem does this paper attempt to address?