Spatially-confined arterial spin-labeling with FAIR.

Yan Zhang,Hee Kwon Song,Jiongjiong Wang,Aranee Techawiboonwong,Felix W Wehrli
DOI: https://doi.org/10.1002/jmri.20362
IF: 4.4
2005-01-01
Journal of Magnetic Resonance Imaging
Abstract:Purpose: To investigate the effectiveness of slab-selective inversion in pulsed arterial spin labeling with body coil excitation as a means to reduce large vessel contamination of the perfusion signal. Materials and Methods: Studies were conducted by varying the tagging width in multislice flow-sensitive alternating inversion recovery (FAIR) in conjunction with body coil excitation on a Siemens Sonata whole-body 1.5-T scanner. The results of spatially-confined tagging were then compared with conventional nonselective tagging in the presence and absence of a bipolar gradient crusher pair in order to determine the effectiveness of suppressing vascular signal and to estimate the bolus width that reaches the capillary bed. Results: It is shown in five volunteers, ages 23-38 years, that depending on the average velocity of the arterial blood flow in the tagging region. a bolus of 6-8 cm in width reaches the capillary bed at a fixed inversion time TI of 1.4 seconds. while a bolus of 11.2-16.5 cm in width enters the imaging region. Further. noticeable velocity differences have been found among the participating subjects, with averages ranging from 10.1 to 13.9 cm/second. Conclusion: The data suggest that it is advantageous to replace nonselective global tagging in FAIR perfusion imaging with body coil excitation by spatially-confined tagging to reduce undesired residual tagged blood in large vessels.