Relationship Between Dosage of SonoVue and Image Quality of Liver Occupied Lesions

杜联芳,苏一巾,李凡,伍瑛
DOI: https://doi.org/10.3969/j.issn.1002-0101.2007.03.019
2007-01-01
Abstract:Objective To explore the relationship between SonoVue dosage and the image quality of liver occupied lesions.Methods Thirty cases with liver occupied lesions and 20 healthy volunteers underwent 2.0 ml,1.2 ml,0.6 ml SonoVue contrast-enhanced ultrasound using Sequoia 512 with probe 4C1-S,imaging technique:CPS;MI:0.21.The injection mode was:5 ml of 0.9% NS solution was injected just after bolus injection of Sonovue via a peripheral vein.Images were recorded by VCD,workstation and analyzed by ACQ software.Results 0.6 ml of SonoVue was suitable for measuring liver artery-vein transit time and displayed clearly less than 1.5 cm liver occupied lesions,2.0 ml SonoVue could show distinctively 5.0 cm lesions,1.2 ml SonoVue could reveal clearly 1.5-5.0 cm lesions.Conclusions Close relationship was present between SonoVue dosage and image quality of liver occupied lesions.