Application of Real-time Three-dimensional Echocardiography in Percutaneous Interventional Procedures for Congenital Heart Disease
Chen Guozhen,Sun Kun,Gao Wei
DOI: https://doi.org/10.3969/j.issn.1002-0101.2007.10.014
2007-01-01
Abstract:Objective To evaluate initial application of RT-3DE in interventional procedures for congenital heart disease(CHD).Methods During the catheter-based interventional procedures,twenty-two patients with CHD underwent RT-3DE examination with Philips iE33 echocardiographic system(Version 2007)and X-matrix transducer.The offline quantitative analyses were made in its Qlab workstation,and their results were compared with those of 2DE.Results The qualitative diagnoses of novel RT-3DE were more informative than those of planar 2DE,objectively demonstrated dynamic detailed visualization of cardiovascular normal anatomy and abnormal pathology such as the anatomic shape,size and position of the septal defect.Furthermore,the whole course of the interventional procedure was guided and monitored online by RT-3DE,providing pushed location of guide-wire and catheter sheath,spatial morphology of device and its relation with contiguous cardiac structures,etc.After the device was first released,there was the incorrect spatial location of Amplatzer occluder in 2 cases with atrial septal defect and the smaller size of device in 1 case with persistent ductus arteriosus.Under the guidance and monitoring of RT-3DE,the second release was correctly completed through replacing the larger device and relocating the position of the delivery sheath and device.In addition,from the quantitative analyses in its Qlab workstation,diameter of septal defect and persistent ductus arteriosus measured by RT-3DE was well correlated with that measured by 2DE(r=0.97),but the measurement of RT-3DE was correlated with the size of its device(r=0.98)better than that of 2DE(r=0.93).Conclusions With instant visualization and online guidance of interventional treatment,RT-3DE will be a safe and valuable imaging technique for selecting preoperative patients,choosing suitable device,performing optimal interventional manipulation and making postoperative effect assessment.