Sensitive Determination of Epinephrine in Pharmaceutical Preparation by Flow Injection Coupled with Chemiluminescence Detection and Mechanism Study.

Yanming Liu,Zhuanli Liu,Yanmei Shi
DOI: https://doi.org/10.1002/bio.1186
IF: 2.613
2009-01-01
Luminescence
Abstract:A novel, rapid and sensitive method was described for the determination of epinephrine (EP) using flow injection analysis coupled with chemiluminescence (CL) detection, which based on EP enhanced the weak CL emission of luminol-KIO(4) system in NaOH solution. Parameters affecting the CL intensity and reproducibility were optimized systematically. Under the optimized experiment conditions, the net CL intensity was proportional to the concentration of EP in the range of 5.0 × 10(-8) to 1.5 × 10(-6)  mol/L with a detection limit of 1.9 × 10(-9)  mol/L. The relative standard deviation (RSD) was found to be 0.7% for 13 replicate determinations of 3.0 × 10(-7)  mol/L EP. The applicability of the proposed method was illustrated in the determination of EP in pharmaceutical preparation. The recoveries of EP at different levels in EP hydrochloride injection were between 95.4 and 104.7%. One assay procedure takes only 27 s, and the sampling rate was calculated about to be 130 samples/h. The possible mechanism of the enhanced CL intensity was studied by examining CL spectra and UV-vis spectra.
What problem does this paper attempt to address?