Preparation of TiO2/PMMA IOL and its safety evaluation by animal experiments

DongSheng Liu,Jingning Weng
IF: 1.645
2008-01-01
International Journal of Ophthalmology
Abstract:• Aim: To prepare intraocular lens with good biocompatibility by TiO2 films surface modified on PMMA intraocular lens. • Methods: Prepare TiO2 films on PMMA substrate with the aid of air plasma treatment to produce a more hydrophilic surface on the PMMA, followed by dipcoating of TiO2 in a TiO2 sol. The safety tests of this material were conducted in Kunming mice and New Zealand rabbits. • Results: After injection and implantation of TiO2/PMMA into mice and rabbits, general condition of animals remained normal. Weight, Hemoglobin, red blood cell count, and white blood cell count of experimental animals altered slightly. Hemolytic rate of TiO2/PMMA was 2.87%, which met the requirement of hemolysis test standard (<5%) as a medical material. Pyrogen test by injecting digestion solution of TiO2/ PMMA in 6 rabbits found a raise in the body Temperature less than 0.6°C, with total increase less than 1.4°C, to meet the accepted criteria for the pyrogen test. Subcutaneous implantation test revealed that implantation did not cause any local redness. There were little inflammatory cell infiltrations around the material 8 weeks after implantation. • Conclusion: TiO2/PMMA intraocular lens have good biocompatibility.