Molecular Sieves Materials Modified Carbon Paste Electrodes for the Determination of Cardiac Troponin I by Anodic Stripping Voltammetry

NY He,HS Guo,SX Ge,D Yang,JN Zhang
DOI: https://doi.org/10.1016/j.micromeso.2005.06.022
IF: 5.876
2005-01-01
Microporous and Mesoporous Materials
Abstract:A novel electrochemical immunoassay for cardiac troponin I (cTnI) combining the concepts of the dual monoclonal antibody "sandwich" principle, the silver enhancement on the nano-gold particle, and the mesoporous materials modified carbon paste electrode (MCPE) is described. Four main steps were carried out to obtain the analytical signal, i.e. electrode preparation, immunoreaction, silver enhancement, and anodic stripping voltammetric detection. The performance of SBA-15 was better than that of MCM-41 probably due to the rather larger pore size of SBA-15. A linear relationship between the anodic stripping peak current and concentration of cTnI from 0.5 to 5.0 ng/ml and a limit of detection of 0.2 ng/ml were obtained using SBA-15 modified CPE (SBA-MCPE), while the linear relationship was from 0.8 to 5.0 ng/ml and the limit of detection was 0.5 ng/ml for MCM-41 modified carbon paste electrode (MCM-MCPE). The established method was tested by determining cTnI in acute myocardial infarction (AMI) samples using enzyme-linked immunoadsorbent assay (ELISA) for comparison analysis, and results from both approaches were aggrement with each other.