Radiofrequency Catheter Ablation of Para-His Accessory Pathway.

XU Jie-wei,LIU Shao-wen,ZHOU Jing-min
2006-01-01
Abstract:Objective To investigate the safety and feasibility of radiofrequency catheter ablation of Para-His accessory pathway.Methods During January,2005 to December,2005,ten patients were diagnosed as Para-His Accessory Pathway(AP) by intracardiac electrophysiologic study.Among them,manifested AP 3 cases,concealed AP 7 cases.The APs were ablated with low power(10w) and in short interval(5s).They employed temperature controlled ablation with energy titration under sinus rhythm.Results Two of ten patients refused ablation because of afraid of high operation risk.All radiofrequency catheter ablations of the eight patients were successful,temperature 55~60℃,power 26.7+11.9w,time 60s,and none was recurred after 3~6 months follow up.Conclusions Radiofrequency catheter ablation of Para-His accessory pathway under sinus rhythm is safe and feasible.But the procedure should be conducted with low power and for a short time at the beginning,and should be under close monitoring.