Abstract:A method for the determination of hemoglobin content in biodegradable polymer microspheres by infrared spectroscopy was developed.By using a appropriate absorption peak and internal standard,the spectral and matrix interferences generated from the internal standard,remnant surfactants and polymer shell were eliminated,thus sample separation was not required prior to analysis.The correlation coefficient of the calibration curve was 0.999 8.The recovery range was 98%-100% for six samples and the relative standard deviation(RSD) was less than 4.2%.The method is rapid and simple.